Field glossary · method
Calibration
Calibration establishes how an instrument's indications relate to reference values, with measurement uncertainty attached to that relationship. It supplies the evidence for interpreting a reading, applying a correction, or deciding whether the instrument meets a tolerance. Adjustment changes the instrument; verification judges it against a requirement. Keeping those operations separate explains why a meter can receive a valid calibration report showing that it failed its accuracy limit.
A scheduled adjustment of the sensor reading against a certified reference gas to correct drift
Specify the calibration deliverable
JCGM 200:2012, the current third edition of the International Vocabulary of Metrology, separates calibration, adjustment, and verification. Its section 3.11 says a measuring system usually needs recalibration after adjustment. Request as-found measurements before settings change and as-left measurements after service; otherwise the report loses the evidence of drift.
A service order for gas-detection wearables or HVAC instruments should name the functions and ranges to be measured, permitted adjustments, and the acceptance specification. The following deliverables distinguish the operations in JCGM's vocabulary.
| Operation | Result to request |
|---|---|
| Calibration | Reference values, instrument indications, and measurement uncertainty at the specified test points. |
| Adjustment | The zero, offset, or span settings changed, followed by new measurements. |
| Verification | A conformity decision against the stated requirement, with the decision rule used. |
Read the evidence behind the calibration sticker
NIST's metrological-traceability policy requires a documented chain of calibrations to a specified reference, with measurement uncertainty contributed at each link. Traceability belongs to the measurement result, rather than to the sticker or the laboratory's name. Obtain the tested function, range, reference values, observed indications, uncertainty, and conditions with the instrument identity and calibration date.
NIST also states that traceability alone does not establish fitness for a measurement task. The uncertainty must suit the decision. A certificate covering DC voltage does not establish the meter's temperature-probe performance. During rooftop sensor work and other BMS point checks, preserve which probe and input channel produced the reported result, particularly when those components are interchangeable.
Read a 33 V result against the correct meter model
Fluke's 80 Series V Calibration Manual, Revision 4 dated November 2025, accepts a Model 87 display of 32.97 to 33.03 V on its 60 V DC range when the calibrator supplies 33 V. At that same test point, Fluke permits 32.96 to 33.04 V for Models 83 and 88. A reading of 33.04 V therefore exceeds the Model 87 limit but falls at the upper limit for the other models. Record the exact model with the result.
At a 33.000 V reference, a Model 87 display of 33.04 V has a +0.040 V error and exceeds its test limit by 0.01 V. Fluke calls for adjustment or repair after a failed performance test. Preserve that as-found result; a later 33.00 V as-left reading documents recovery at this point, not acceptance of the meter's other ranges and functions. Use the failed function and range to identify earlier job readings for review, and include the laboratory's uncertainty and decision rule with its conformity statement.
Set the next interval from operating history
Fluke's November 2025 calibration manual calls for annual 80 Series V performance tests and states its accuracy specifications for one year after adjustment at 18 to 28 degrees C and up to 90% relative humidity. That is an equipment-specific interval and reference environment. The temperature range describes the accuracy conditions, rather than the meter's full operating-temperature rating.
NIST does not prescribe a general recalibration period; it identifies required accuracy, contract or regulatory requirements, instrument stability, and environment as interval-setting factors. ILAC-G24:2022/OIML D 10:2022 recommends retaining calibration history to support later interval decisions and maintaining intermediate checks between calibrations. Track the as-found error against the permitted error at successive returns. That history shows whether the instrument is holding its limit through the assigned interval.
Distinctions
- Bump test
- A bump-test result records functional response. A calibration result reports the relationship needed to interpret a numerical indication and its uncertainty.
- Zero adjustment
- JCGM 200:2012 treats zero adjustment as an adjustment operation. Setting the zero does not establish gain or linearity across the range; the calibration report needs the reference comparisons for the points being claimed.
Frequently asked questions
Can an instrument be calibrated and still fail?
Yes. Calibration can establish an error outside the permitted limit. An adjustment or repair, followed by new measurements, is a separate action. Keep both results so the report explains what arrived and what left.
Is 'NIST certified' the right description for a field meter?
Use the actual calibration provider and the report's traceability statement. NIST says its calibration results do not constitute product approval or a guarantee of later performance.
Sources checked
- BIPM JCGM publications: JCGM 200:2012 and annotated VIM3
- JCGM 200:2012, annotated VIM3, 2.39 Calibration
- JCGM 200:2012, annotated VIM3, 3.11 Adjustment of a measuring system
- JCGM 200:2012, annotated VIM3, 2.44 Verification
- NIST: Metrological Traceability, policy and frequently asked questions
- Fluke: 80 Series V Calibration Manual, Revision 4, November 2025, Performance Tests and Table 3
- NIST: Recommended Calibration Interval, updated May 29, 2026
- ILAC-G24:2022/OIML D 10:2022: Guidelines for the determination of recalibration intervals, sections 4.8 and 4.9
Also defined as
- HVAC: Checking an instrument against a traceable reference on an interval, plus the field zero or fresh-air check before reading